Hardware Specifications Base Package


Use of standard PC technology instead of exotic and proprietary DSP hardware delivers the highest computing performance and guarantees worldwide availability now and in the long term. The National Instruments Real-Time operating system delivers the highest connectivity with drivers for almost any interface including USB, serial, GPIB and many more. LabVIEW Real-Time and LabVIEW FPGA is a high-level programming language and will be portable to latest hardware developments, thus making the Nanonis controller also future-proof. Upgrade path becomes much more straightforward and cost effective.

  Realtime Controller RC4
  • dimensions, weight: 32.5 x 28 x 12 cm, 4.5 kg
  • power: universal power supply, max. 60 W
  • components: Pentium mobile 2 GHz, 1 GB ram, 40 GB HD, 2 PCI slots, ethernet adapter
  • operating system: National Instruments Real-time OS
  • I/O card: NI-FPGA RIO (PCI-7831R)

  Signal Conditioning SC4
  • dimensions, weight: 32.5 x 28 x 7 cm, 3.6 kg
  • power: 100 V - 240 V, 50 - 60 Hz, 10 W, automatic switching, toroidal transformer, linear regulated
  • ground: 100 kΩ AGND to chassis, decoupled from RC4

  Analog Inputs
  • hardware interface: 8 x BNC connectors, differential
  • diff. input volt. range: ±10 V
  • diff. input resistance: 100 kΩ
  • analog bandwidth: DC - 5 kHz, 4th order Butterworth low-pass filter
  • AD converter: 16-bit, no missing codes, 200 kS/s
  • effective resolution: 20-bit @ 10 kS/s, 24-bit @ 100 S/s (oversampling)
  • analog input noise: < 15 nV/sqrt[Hz]
  • converter noise: < 2 nV/sqrt[Hz]
  • measurement noise: < 140 µV rms @ 10 kS/s, < 45 µV rms @ 1 kS/s

  Analog Outputs
  • hardware interface: 8 x BNC, referenced to AGND
  • output voltage: ±10 V into 2 kΩ, (0 to 10 V per dip switch per channel)
  • output resistance: < 1 Ω, short circuit safe
  • analog bandwidth: 5 kHz, 4th order Butterworth low-pass filter
  • DA converter: 16-bit, monotonic, 1 MS/s
  • effective resolution: 22-bit, patented hrDACTM technology with active glitch compensation
  • noise density: < 350 nV/sqrt[Hz]
  • output noise: < 30 µV rms, < 150  µV pp (DC - 1 MHz)

   Digital Lines and Pulse Counter
  • ports: 2 x 8 lines on two sub-D9 female
  • direction: input or output for each line
  • signal: 3.3 V TTL, max. 5 mA per line
  • pulse counter: 2, up to 40 MC/s

  Coarse Positioning and Auto Approach
  • types: slip stick piezo drive, beetle type, stepper motor, DC motor
  • interfaces: digital port, serial, USB, GBIP, ethernet, etc.
  • auto approach: different protocols, configurable, programmable with programming interface

 


Back  |   Print Page  |   Recommend Page

nanonis   News & Events
Upgrade Kit for SPM100



Upgrade your controller to a fully digital system.
More information »


9.10.2008
Adaptation Kit for SPECS STMs



The Adaptation Kit for SPECS STMs further extends the list of commercial microscopes that are supported by the Nanonis controller.

All News »

nanonis   Address


             Nanonis GmbH
             Technoparkstrasse 1
             8005 Zurich
             Switzerland

             Tel.:    +41-44 445 18 25
             Fax:    +41-44 445 18 26
             Email: info@nanonis.com

© 2008 by Nanonis Powered by MySign